LogTag® UTRID-16

Record temperature and humidity at little cost  Reliable, uncomplicated, credit card sized  Unbeaten value for money

The LogTag ® UTRID-16 provides a display to allow immediate viewing of current and past temperature and alarm statistics. The display shows recording status, current temperature, alarm status of up to six user-configurable alarms and duration of any temperature excursions. A detailed PDF report of the recorded temperature data and alarms are automatically generated without any proprietary software using the integrated USB plug

  • Records temperature from -30°C to +60°C

  • UTRID-16 is Pre-Qualified compliant to WHO PQS specification E006 Temperature monitoring devices

  • Up to 16,000 recordings - enough for 168 days at 15 minute logging.

  • Can be connected to any computer with a USB port. No interface cradle required.

  • Display showing temperature and alarm status of up to six alarms.

  • UTRID-16 offers both a user Replaceable battery and a Fixed Battery type

  • Automatic generation of a PDF report as a permanent record of the data.

  • Creates a comprehensive data file compatible with LogTag Analyzer operating software.

     

    LogTag UTRID-16

    Sensor Measurement Range

    -30°C to +60°C (-22°F to +140°F).

    Operating Temperature Range

    -30°C to +65°C (-22°F to +149°F)

    Storage Temperature Range

    0°C to +40°C (32°F to +104°F)

    Rated Temperature Reading Resolution

    0.1°C (0.1°F)

    Rated Temperature Reading Accuracy

    Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F).
    Better than ±0.7°C (±1.3°F) for -25°C to -20°C (-13°F to -4°F) & +40°C to +60°C (+104°F to +140°F)

    Sensor Reaction Time

    Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999

    Recording Capacity

    16,129 temperature readings.
    112 days @ 10min logging, 168 days @ 15min logging. Statistics memory (for display on LCD): Max/Min values and alarm duration values

    Sampling Interval

    Configurable from 30 seconds to 18 hours

    Logging Start Options

    Push button start or specific date & time. Optional start delay of up to 72 hours

    Recording Indication

    “REC” symbol on LCD

    Alarms

    Three configurable upper and three configurable lower alarms

    Download Time

    Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report.
    Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured)

    Environmental

    IEC 60529: IP64 with USB cap fitted.

    Power Source

    CR2450 3V LiMnO2
    UN38.3 approved coin cell battery (Fixed)

    Battery Life

    Typically 2 years of operation with normal use (6 minute logging, statistics reviewed on the display no more than once daily for no longer than 30
    seconds each time, download data monthly) provided recorder is kept within the storage temperature range when not in use

    Connection Interface

    USB 2.0, A-type plug

    Software

    PDF Reader, LogTag® Analyzer 2.8 or higher

    PDF Report

    One page report with trip and alarm summary. Multi-page report with list of readings, including date and time

    Size

    93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap.

    Weight

    39g

    Case Material

    Polycarbonate.